Fluke Calibration 4180 [4180-256] IR Calibrator -15 to 120°C

Not yet rated Write a review
Fluke Calibration
Rp.192.500.000
Availability:
4 - 6 WEEKS
Product Code:
4180-256
Brand:
Fluke Calibration
Update Terakhir:
13 May 2017
  • Free delivery

  • Money back guarantee

Share

The Fluke Calibration 4180256 Precision Infrared (IR) Calibrator, -15 °C to 120 °C (5 °F to 248 °F) for infrared thermometers and thermal imagers is fast, accurate, and easy to use. It comes with an accredited calibration, sample calibration procedures for Fluke thermometers built right in, and everything you need to start making high-quality infrared thermometer calibrations. The 4180-156 is the perfect solution for any infrared thermometer or thermal imager within its temperature range.

The Fluke Calibration 4180-256 reaches temperatures from -15 °C to 120 °C. Uniformity is important in infrared temperature calibration work because an infrared thermometer will "see" as much as the entire target when placed at the appropriate calibration distance and each pixel of a thermal imager registers a temperature that needs to be both accurate and consistent across the imager.

Lastly, the 4180-256 can meet its specifications without additional emissivity related corrections, leading to legitimate test uncertainty ratios (TUR) as good as 4:1.

  • Calibration solutions from –15°C to 500°C (5°F to 932°F)
  • Display accuracy up to ±0.35°C
  • Stability up to ±0.05°C
  • Uniformity up to ±0.10°C
  • Rapid heating and cooling
  • RS-232 interface capability
  • Correct target size for most thermometers
  • Radiometrically calibrated for traceable and consistent results

  • Temperature range (@ 23 °C ambient, 0.95 emissivity): –15 °C to 120 °C
Display accuracy:
  • ± 0.40 °C at –15 °C
  • ± 0.40 °C at 0 °C
  • ± 0.50 °C at 50 °C
  • ± 0.50 °C at 100 °C
  • ± 0.55 °C at 120 °C
Stability:
  • ± 0.10 °C at –15 °C
  • ± 0.05 °C at 0 °C
  • ± 0.10 °C at 120 °C
Uniformity (5.0 in dia of center of target):
  • ± 0.15 °C at –15 °C
  • ± 0.10 °C at 0 °C
  • ± 0.25 °C at 120 °C
  • Uniformity (2.0 in dia of center of target):
  • ± 0.10 °C at -15 °C
  • ± 0.10 °C at 0 °C
  • ± 0.20 °C at 120 °C
  • Heating time:
  • 15 min: –15 °C to 120 °C
  • 14 min: 23 °C to 120 °C
  • Cooling time 15 min:
  • 120 °C to 23 °C
  • 20 min: 23 °C to –15 °C
  • Stabilization time: 10 minutes
  • Nominal emissivity:  0.95
  • Thermometer emissivity compensation: 0.9 to 1.0
  • Target diameter: 0152.4 mm (6 in)
  • Computer interface: RS-232
  • Power: 15 V ac (± 10%), 6.3 A, 50/60 Hz, 630 W. 230 V ac (± 10%), 3.15 A, 50/60 Hz, 630 W
  • Fuse(s): 115 V ac 6.3 A, 250 V, slow blow. 230 V ac 3.15 A, 250 V, T
  • Size (HxWxD): 356 mm x 241 mm x 216 mm (14 in x 9.5 in x 8.5 in)
  • Weight: 9.1 kg (20 lb)
  • Safety: EN 61010-1:2001, CAN/CSA C22.2 No. 61010.1-04

  • (1) 4180 IR Calibrator
  • (1) Report of Calibration (NVLAP Accredited) Certificate
  • (1) Power Cord
  • (1) User’s Guide
  • (1) Getting Started Guide
  • (1) Documentation CD
  • (1) Target Cover
  • (1) Serial Cable

Download