• Vaisala HMP3 [HMP3-A-2-A-1-A-0-A-0-0-2] General Purpose Humidity and Temperature Probe w/2m Cable, Humicap R2 sensor, No Sensor Purge, 19200 Baud Rate, Plastic Grid & SST Net Filter & 10m Cable Connection with Open Wire

Vaisala HMP3 [HMP3-A-2-A-1-A-0-A-0-0-2] General Purpose Humidity and Temperature Probe w/2m Cable, Humicap R2 sensor, No Sensor Purge, 19200 Baud Rate, Plastic Grid & SST Net Filter & 10m Cable Connection with Open Wire

Not yet rated Write a review
Vaisala
Rp.51,111,200
Availability:
4 - 6 WEEKS
Product Code:
HMP3-A-2-A-1-A-0-A-0-0-2
Brand:
Vaisala
Update Terakhir:
18 Jun 2024
  • Free delivery

  • Money back guarantee

Ask an Expert

Our tech experts are ready to help!

(62) 021-53167044 | 0812-83861970 | 0812-9915277

Email Us

Share

Vaisala HMP3 [HMP3-A-2-A-1-A-0-A-0-0-2] General Purpose Humidity and Temperature Probe w/2m Cable, Humicap R2 sensor, No Sensor Purge, 19200  Baud Rate, Plastic Grid & SST Net Filter & 10m Cable Connection with Open Wire. Vaisala HUMICAP Humidity and Temperature Probe HMP3 is a general-purpose probe designed for various industrial processes. The probe structure allows for replacing the sensor without tools, making the probe suitable for applications such as paint booths and other industrial applications where periodic recalibration alone is not su€cient for maintaining the probe performance. Other applications include, for example, industrial HVAC systems, cleanrooms, and environmental chambers.

Probe design allows for several operating environments and flexible field maintenance. Filter and HUMICAP R2 sensor element are field replaceable for applications that require frequent replacements. Calibration and adjustment of humidity measurement is also needed if the HUMICAPâ R2 sensor is replaced.

If purchased with a composite sensor instead of the field-replaceable HUMICAPâ R2 sensor, HMP3 can use the sensor purge feature. In environments with high concentrations of chemicals and cleaning agents, sensor purge helps to maintain measurement accuracy between calibration intervals. Sensor purge involves heating the sensor to remove harmful chemicals. The function can be initiated manually or programmed to occur at set intervals.

The probe can be used as a standalone digital Modbus RTU transmitter over an RS-485 serial bus, and it can also be connected to Indigo transmitters and the Indigo80 handheld indicator. For easyto-use access to field calibration, device analytics, and configuration functionality, the probe can be connected to Vaisala Insight software for Windows.

Indigo transmitters extend the capabilities of Indigo-compatible measurement probes. The transmitters can display measurements on the spot as well as transmit them to automation systems through analog signals, digital outputs, and relays. Cable length between probe and transmitter can be extended to up to 30 meters. The Indigo80 handheld indicator is ideal for spot-checking and process monitoring, as well as for configuring, troubleshooting, calibrating, and adjusting the probe.

There are no reviews for this product.

Write a review

  • Part Number: HMP3-A-2-A-1-A-0-A-0-0-2
  • Probe Head: HMP3 general purpose probe
  • Cable length: HMP3 with 2m cable
  • Sensor Type: Humicap R2 sensor, no sensor purge
  • Filter Type: Plastic grid & SST net filter, HMP3, HMP7
  • Purge and Heating: Purge off, probe heat off
  • Baud Rate: 19200 - Use with Indigo transmitters1
  • Parity, Data Bit, Stop Bit:  8, N, 2 - Use with Indigo transmitters1
  • Probe Modbus Address: 240 - Use with Indigo transmitters1
  • Certificate: English certificate
  • Probe Mounting Accessory: No accessories
  • Connection Cable: 10m connection cable with open wires
  • Available with field-replaceable
  • HUMICAP R2 sensor
  • RH accuracy up to 0.8 %RH
  • Temperature accuracy up to 0.1 °C (0.18 °F)
  • Temperature measurement range −40 … +120 °C (−40 … +248 °F)
  • Sensor purge improves long-term stability and chemical resistance
  • Modbus RTU over RS-485
  • Compatible with Vaisala Indigo products and Insight PC software

  • Relative Humidity:
  • Measurement range: 0 … 100 %RH
  • Accuracy at +23 °C (+73.4 °F) 1):  ±0.8 %RH (0 … 90 %RH)
  • Factory calibration uncertainty 2):  ±0.5 %RH (0 … 40 %RH); ±0.8 %RH (40 … 95 %RH)
  • T63 response time: 15 s
  • Sensor options: HUMICAP R2; HUMICAPâ R2C 3); HUMICAPâ 180VC 3) 4)
  • Temperature:
  • Sensor: Pt100 RTD Class F0.1 IEC 60751
  • Measurement range:  −40 … +120 °C (−40 … +248 °F)
  • Accuracy 1):  ±0.1 °C (±0.18 °F)
  • Factory calibration uncertainty 2):  ±0.1 °C (±0.18 °F) at +23 °C (+73.4 °F)
  • 1) Defined against calibration reference. Including non-linearity, hysteresis, and repeatability.
  • 2) Defined as ±2 standard deviation limits. Small variations possible; see calibration certificate.
  • 3) Sensor purge feature available with this sensor.
  • 4) H2O2 resistant. With HUMICAPâ 180VC sensor, accuracy is not specified below −20 °C (−4 °F) operating temperature.
  • Operating environment:
  • Operating temperature of probe head: −40 … +120 °C (−40 … +248 °F)
  • Operating temperature of probe body: −40 … +80 °C (−40 … +176 °F)
  • Storage temperature : −40 … +80 °C (−40 … +176 °F)
  • Operating environment: Suitable for outdoor use
  • Measurement environment: For air, nitrogen, hydrogen, argon,helium, and oxygen 1)
  • IP rating of probe body:  IP66
  • 1) Consult Vaisala if other chemicals are present. Consider safety regulations with flammable gases.
  • Inputs and outputs:
  • Operating voltage: 15 … 30 V DC
  • Current consumption: 10 mA typical, 500 mA max.
  • Digital output: RS-485, non-isolated
  • Protocols: Modbus RTU
  • Output parameters:
  • Absolute humidity (g/m3): Relative humidity (%RH)
  • Absolute humidity at NTP (g/m3): Relative humidity (dew/frost) (%RH)
  • Dew point temperature (°C): Temperature (°C)
  • Dew/frost point temperature (°C): Water concentration (ppmv)
  • Dew/frost point temperature at 1 atm (°C): Water concentration (wet basis) (vol‑%)
  • Dew point temperature at 1 atm (°C): Water mass fraction (ppmw)
  • Dew point temperature diference (°C): Water vapor pressure (hPa)
  • Enthalpy (kJ/kg); Water vapor saturation pressure (hPa)
  • Mixing ratio (g/kg): Wet-bulb temperature (°C)
  • Compliance:
  • EU directives and regulations: EMC Directive (2014/30/EU); RoHS Directive (2011/65/EU) amended by 2015/863
  • EMC compatibility: EN 61326‑1, industrial environment
  • Type approvals: DNV GL certificate no. TAA00002YT
  • Compliance marks: CE, China RoHS, RCM
  • Mechanical specifications:
  • Connector: M12 5-pin A-coded male
  • Weight: 302 g (10.65 oz)
  • Probe cable length: 2 m (6.56 ft)
  • Materials:
  • Probe: AISI 316L
  • Probe body: AISI 316L
  • Cable jacket: FEP

  • (1) Vaisala HMP3 General Purpose Humidity and Temperature Probe
  • (1) Instruction Manual

Download

Note: Prices are not binding at any time subject to change without prior notice