• Ametek RTC-159B Reference Temperature Calibrator, -100˚C to 155˚C (-148˚F to 311˚F), Full Model

Ametek RTC-159B Reference Temperature Calibrator, -100˚C to 155˚C (-148˚F to 311˚F), Full Model

Not yet rated Write a review
Ametek

Available Options

Rp.918,295,200
Availability:
4 - 6 WEEKS
Product Code:
RTC159B
Brand:
Ametek
Update Terakhir:
16 May 2024
  • Free delivery

  • Money back guarantee

Share

Ametek RTC-159 Reference Temperature Calibrator,  -100˚C to 155˚C (-148˚F to 311˚F), Full Model. The Ametek RTC-159 Reference Temperature Calibrator is the most versatile temperature calibrator available with a temperature range that makes it especially ideal for use in the health care, medical, pharmaceutical, biotechnology and food industries. The RTC-159 can be used in many applications where either high heat or extreme cooling is needed. Intuitive to use and easy to run, the RTC-159 is equipped with a large informative, easy-to-read color display. The RTC-159 is lightweight and easy-to-carry and ensures durability. The RTC-series features our Ametek's active dualzone heating technology. Each heating zone is independently controlled for precision temperature calibration. The homogeneity in the lower part is close to that of a laboratory liquid bath. The lower zone ensures optimum heat dissipation throughout the entire calibration zone. The upper zone compensates for heat loss from the sensor-under-test and from the open top. This design also eliminates the need for extra insulation of sensors-under-test and makes it possible to calibrate any type of mechanical sensors. The RTC-159 can perform calibration over a very wide temperature range starting from -100°C and up to 155°C (-148 to 311°F). The patented technology makes it possible to perform calibration of sensors in applications ranging from ultra-coolers to autoclaves.

  • Full model; w/ DLC, reference, under-test inputs
  • Temperature range, -100˚C to 155˚C (-148˚F to 311˚F)
  • Best-in-class temperature homogeneity
  • Widest temperature range
  • Fastest temperature calibration
  • Intelligent reference sensor communication
  • Intelligent recalibration information (IRI)
  • USB connector for communication
  • Efficient cooling technology
  • Easy to read & user friendly
  • Easy to carry
  • SET-Follows-TRUE (Models B & C only)
  • Reading of sensor-under-test (Model B only)
  • Switch test (Model B only)
  • Auto stepping
  • Instrument setups
  • Maximum and minimum temperatures
  • As Found/As Left (Model B only)
  • Calibration of indication devices
  • Enhanced stability
  • Specially-designed carrying case

  • Temperature range:
  • @ ambient temp. 0°C/32°F: -100 to 155°C/-148 to 311°F
  • @ ambient temp. 23°C/73°F: -100 to 155°C/-148 to 311°F
  • @ ambient temp. 40°C/104°F: -83 to 155°C/-117 to 311°F
  • Accuracy (model B & C) with external STS ref. sensor: 159 B & C:  ±0.06°C/±0.11°F
  • 12-month period. Relative to reference standard. Specifications by use of the external JOFRA STS-200 reference sensor
  • Accuracy with internal reference sensor:
  • RTC-159 A, B & C:  ±0.30°C/±0.54°F
  • Stability:  ±0.03°C/±0.054°F1)
  • Measured after the stability indicator has been on for 15 minutes. Measuring time is 30 minutes.
  • Radial homogeneity (difference between holes):
  • RTC-159: 0.01°C/0.02°F
  • Resolution (user-selectable):
  • All temperatures: 1° or 0.1° or 0.01° or 0.001˚
  • Heating time:
  • RTC-159 -100 to 23°C/-148 to 73°F:  12 minutes
  • 23 to 155°C/73 to 311°F: 14 minutes
  • Cooling time:
  • RTC-159 155 to 23°C/311 to 73°F:  40 minutes
  • 23 to -80°C/73 to -112°F:  65 minutes
  • -80 to -90°C/-176 to -130°F: 20 minutes
  • -90 to -100°C/-194 to -148°F:  50 minutes
  • 23 to -100°C/73 to -148°F: 135 minutes
  • 155 to -100°C/311 to -148°F:  175 minutes
  • Time to stability (approx.):
  • RTC-159: 10 minutes
  • Immersion depth:
  • RTC-159:  190 mm/6.3 in
  • Thermocouple input:
  • Thermocouple types: E, J, L, K, N, R, S, T, U, B
  • Range : ±78 mV
  • F.S. (Full Scale) : 78 mV
  • Accuracy (12 months) : ±(0.005% Rdg. + 0.005% F.S.)
  • Transmitter supply:
  • Output voltage: .... 24VDC ±10%
  • Output current: Maximum 28 mA
  • Transmitter input mA (B model only):
  • Range: 0 to 24 mA
  • Accuracy (12 months) : ±(0.005% Rdg. +0.010% F.S.)
  • Voltage input VDC (B model only):
  • Range:  0 to 12 VDC
  • Accuracy (12 months) : ±(0.005% Rdg. +0.010% F.S.)
  • Switch input (B model only):
  • Switch dry contacts
  • Test voltage: Maximum 5 VDC
  • Test current : Maximum 2.5 mA
  • Mains specifications:
  • Voltage : 115V (90-127) / 230V (180-254)
  • Frequency, non US deliveries : 50 Hz ±5, 60 Hz ±5
  • Frequency, US deliveries : 60 Hz ±5
  • Power consumption (max.) : 450 VA
  • RTD reference input (B & C models only):
  • Type:  4-wire RTD with true ohm measurements1)
  • F.S. (Full Scale) : 400 ohm
  • Accuracy (12 months) :  ±(0.0012% rdg. + 0.0005% F.S.
  • Communication interface:
  • Serial data interface: USB 2.0 device port
  • Serial data interface : USB 2.0 host double port*
  • LAN:  Ethernet MAC 10/100 Base-T*
  • SD: SD slot*
  • * for future expansion
  • Miscellaneous:
  • Operating temperature: 0 to 40°C/32 to 104°F
  • Storage temperature : -20 to 50˚C/-4 to 122˚F
  • Humidity : 0 to 90% RH
  • Protection class : IP-10
  • PHYSICAL SPECIFICATIONS:
  • Weight and instrument size (L x W x H):
  • RTC-159: 531 x 169 x 432 mm/20.9 x 6.65 x 17.0 in
  • Shipping (including carrying case):
  • RTC-159: 38 kg/83.8 lb
  • RTC-159: 800 x 500 x 800 mm/31.5 x 19.7 x 31.5 in
  • Note: Shipped on ½ pallet, binded
  • INSERTS:
  • Insert dimensions:
  • RTC-159 outer diameter: 29.7 mm/1.17 in
  • RTC-159 inner diameter: 25.6 mm/1.01 in
  • RTC-159 length: 150 mm/5.91 in
  • Weight of non-drilled insert (approx.):
  • RTC-159: 290 g/10.2 oz
  • RTC-159: 15.2 kg/33.5 lb
  • Use of other inserts may reduce the performance of the calibrator. To get the best results, the insert dimensions, tolerance and material is critical. We advise using JOFRA inserts, as they guarantee                   trouble-free operation

  • (1) RTC dry-block calibrator with DLC, reference, under-test inputs
  • (1) Mains power cable
  • (1) Traceable certificate - temperature performance
  • (1) Traceable certificate - input performance for reference sensor and DLC sensor
  • (1)  Traceable certificate - input performance for sensor-under-test inputs
  • (1) Test cables (2 x red, 2 x black)
  • (1) Tool for insertion tubes
  • (1) JOFRACAL
  • (1) USB cable
  • (1) Set of rubber cones for insulation plugs
  • (1) Carrying case
  • (1) Instruction Manual

          Download

          Note: Prices are not binding at any time subject to change without prior notice