• Ametek CTC-350A [CTC350A230BM01XXFCX] Basic Temperature Calibrator 230V EU 28 to 350°C (82 to 662°F) w/ Multi-Hole Insert (Metric) and Solid Protective Carrying Case

Ametek CTC-350A [CTC350A230BM01XXFCX] Basic Temperature Calibrator 230V EU 28 to 350°C (82 to 662°F) w/ Multi-Hole Insert (Metric) and Solid Protective Carrying Case

Not yet rated Write a review
Ametek
Rp.141.500.000
Availability:
4 - 6 WEEKS
Product Code:
CTC350A230BM01XXFCX
Brand:
Ametek
Update Terakhir:
13 May 2017
  • Free delivery

  • Money back guarantee

Share

The Ametek CTC Series is a fast, time-saving, reliable dry block temperature calibrator designed for on-site use. Available in three models with ranges from -25 to 155°C (-13 to 311°F), 28 to 350°C (82 to 662°F), and 28 to 660°C (82 to 1220°F); the CTC series offers the speed, portability, cleanliness, and wide temperature range which are all superior to liquid bath systems.

The CTC series provides exceptional accuracy and stability. The large color display is easy-to-read and provides icons and information regarding the status of the calibrator and the calibration in process. Popular functions like automatic switch test and auto stepping are available with special one-key-one-function buttons. When switching on the calibrator or connecting the reference sensor, the calibrator immediately warns you if any of the calibration certificates are overdue. A buzzer and warning appears. The recalibration interval can be set from 1 to 99 months.

The CTC series is available with an optional signal input for an external reference sensor. The sensor makes it possible to improve accuracy even more. The CTC can be run in two modes when using the external reference sensor. "External ref" mode is where the external reference sensor represents the True value. The "Set follows True" mode is where the reference sensor serves two purposes; measuring the reference temperature and at the same time controlling the block temperature to the set temperature.

The Ametek CTC is supplied with our highly versatile calibration software JofraCal. All calibrations can be documented with a certificate, given that the CTC is controlled from a PC. When the calibrator has reached the desired temperature and stabillity it will prompt you to type in the UUT temperature. JofraCal documents all your calibration needs within temperature, pressure and process calibration.

  • Part Number: CTC350A230BM01XXFCX
  • Temperature Range: 28 to 350°C (82 to 662°F)
  • Model: Basic model, without input
  • Power Supply: 230 VAC EU Version
  • Insert Type: Metric sizes
  • Insert Size: Multi-Hole (3mm, 2 x 4mm, 5mm, 6mm, 9mm)
  • External Reference Sensor: Without
  • Calibration Certificate: Traceable to International Standards
  • Accessories: Solid protective carrying case

  • Temperature: 
  • Temp. @ ambient 23°C / 73° F: -25 to 155°C /-13 to 311°F
  • Temp. @ ambient of 0°C / 32° F:  -39 to 155°C / -38 to 311°F
  • Temp. @ ambient of 50°C / 122° F:  -7 to 155°C / -19 to 311°F
  • Accuracy: 
  • CTC-155 with internal ref. sensor: ±0.3°C /±0.54°F
  • CTC-155 with ext. ref. sensor STS-102-A: ±0.2°C /±0.36°F
  • CTC-155 with ext. ref. sensor STS-120-A-915: ±0.2°C /±0.36°F
  • Stability:
  • CTC-350:  ±0.05°C / ±0.09°F
  • Radial Homogeneity (difference between holes):
  • CTC-155 @ 155°C / 311°F:  0.03°C / 0.054°F
  • CTC-155 @ -25°C / -13°F:  0.02°C / 0.036°F
  • Influence From Load Ø6mm 350°C / 662°F:
  • With Internal Reference:  0,2°C / 0.36°F
  • With External Reference:  0,01°C / 0.02°F
  • Settings: 
  • Resolution: 1 or 0.1 or 0.01
  • Units: °C or °F or K
  • Heating Time:  
  • CTC-350:  23 to 350°C / 73 to 662°F:  6 minutes
  • Cooling Time:
  • CTC-350: 350 to 100°C / 662 to 212°F: 20 minutes
  • CTC-350: 100 to 50°C / 212 to 122°F: 14 minutes
  • Time to Stability (typical):
  • CTC-350:  10 minutes
  • Mains Power: 
  • Voltage: 115 V (90-127) / 230 V (180-254)
  • Max Power Consumption:  130 VA
  • Frequency, US deliveries: 60 Hz ±3
  • Frequency, non US deliveries: 50 Hz ±3, 60 Hz ±3
  • Physical Specifications:
  • Dimension:  L x W x H: 248x148x305 mm / 9.76x5.83x12.01 in
  • Weight: CTC-350:  5 kg / 11 lb
  • Immersion Depth: CTC-350: 115 mm / 4.53 in
  • Well Diameter: CTC-350: 26 mm / 1.02 in
  • Insert Dimensions (diameter x length): CTC-350: 25.7 mm x 120 mm / 1.01 x 4.72 in
  • Electrical: 
  • Switch Input (dry contact)
  • Test Voltage: Maximum 14 VDC
  • Test Current: Maximum 1 mA
  • Digital Interface: USB 2.0
  • Environmental:
  • Operating Temperature: 0 to 50°C / 32 to 122°F
  • Storage Temperature: -20 to 50°C / -4 to 122°F
  • Humidity: 5 to 90% Rh, non-condensing
  • Protection Class: IP-10
  • External Reference Sensor:
  • STS-120-A-915:  -25 to 155°C /-13 to 311°F
  • Accuracy:
  • Hysteresis (@ 0°C / 32°F): 0.01°C / 0.018°F
  • Long Term Stability (@ 0°C / 32°F) : 0.014°C / 0.025°F
  • Repeatability.:  0.004°C/ 0.007°F
  • Sensing Element: 
  • Type:  Pt100
  • Response Time: 
  • STS-120-A: t: 0.5 (50%): 7 sec.
  • STS-120-A: t: 0.9 (90%): 18 sec.
  • Dimensions:
  • Diameter: 4 mm / 0.157 in
  • Length: 140 mm / 5.51 in
  • Max. height over calibrator top:  20 mm / 0.79 in
  • External Reference Sensor: 
  • STS-102-A:  -50 to 155°C /-58 to 311°F
  • Accuracy: 
  • Hysteresis (@ 0°C / 32°F):  0.01°C / 0.018°F
  • Long Term Stability (@ 0°C / 32°F): 0.014°C / 0.025°F
  • Repeatability: 0.002°C/ 0.0036°F
  • Sensing Element: 
  • Type:  Pt100
  • Response Time: 
  • STS-120-A: t: 0.5 (50%): 5 sec.
  • STS-120-A: t: 0.9 (90%): 16 sec.
  • Dimensions: 
  • Diameter: 4 mm / 0.157 in
  • Length: 30 mm / 1.181 in
  • Cable length: 1 m / 3.28 ft

  • (1) CTC dry-block calibrator (user specified)
  • (1) Mains power cable (user specified)
  • (1) Traceable certificate - temperature performance
  • (1) Tool for insertion tubes
  • (1) User manual
  • (1) Test cables (1 x red, 1 x black)
  • (1) USB cable
  • (1) JofraCal calibration software
  • (1) Metric Insert Multi-Hole (3mm, 2 x 4mm, 5mm, 6mm, 9mm)
  • (1) Solid protective carrying case

Download

Note: Prices are not binding at any time subject to change without prior notice