The Chauvin Arnoux MX1 Multimeter with Carrying Case is accompanied with a built-in safety alarm that activates when the Ohmmeter is being used. On the front of the meter there is also a light which will flash to you when a fuse has been blown inside. 

A lot of analogue multimeters on the market are not able to stand much more than a steady level of voltage, but this is where the Metrix MX1 is different. Even if the level of Vrmns reaches as high as 400, no damage will be done. The Ohmmeter function is even completely protected by an electronic circuit. And because this circuit doesn’t have a fuse built in, it becomes more reliable and more easy to use. 

The input terminals that are included in this analogue multimeter can take transient pulses that measure up to 6kV. The current ranges are also fully protected by the inclusion of several HRC (high rupturing capacity) safety fuses that protect your device up to 600V.

When you use the continuity test, that also has an audible signal connection ‘beep’ in order to check if your product is carrying out standard resistance measurements.This particular analogue multimeter also comes with its own carrying case. This case has been specially designed to be completely waterproof and shock proof, so it can withstand a few knocks and drops. The casing itself is also very smooth, making it simple to clean.

The Chauvin Arnoux MX1 Multimeter with Carrying Case is doubly insulated, which ensures that the leads can withstand levels of voltage up to 1500V.

  • 9 DC voltage ranges: 150mV to 1500V
  • 6 AC voltages ranges: 5V to 1500V
  • 7 DC current ranges: 50μA to 10A
  • 6 AC current ranges: 500mA to 10A
  • 3Ω ranges to 20kΩ
  • -6 to +15dB scale for AC measurements
  • Diode test capability with Ωx1 range
  • Mains proof ohms range
  • 10kA fuse protection on current ranges
  • High speed continuity bleeper (<150Ω)
  • Audible warning if >5V is applied to Ω range
  • Visual warning if protection fuses blown
  • Locking system prevents test leads falling out
  • IP65 rated case which cannot be opened with test leads connected, plus moulded-on shock absorber
  • Safety Rating: EN61010-1 CAT II 600 V

  • DC VOLTAGES:
  • Calibres:  150 mV - 0,5 V - 1,5 V - 5 V - 15 V - 50 V - 150 V - 500 V - 1,5 kV (1)
  • Protection: 320 Veff (cal. ≤ 5 V) - 500 Veff (15 V); 1000 Veff (50 V) - 1500 Veff (150 V to 1,5 kV)
  • Accuracy class: 2
  • Input impedance: 20 kΩ /V
  • AC VOLTAGES:
  • Calibres: 5 V - 15 V - 50 V - 150 V - 500 V - 1,5 kV (1)
  • Protection: 320 Veff (cal. ≤ 15 V) - 500 Veff (50 V); 1000 Veff (150 V) - 1500 Veff (5000 V and 1,5 kV)
  • Bandwidth: 16 Hz à 1 kHz (500 Hz / 1500 V)
  • Accuracy class: 2,5
  • Input impedance: 6,32 kΩ /V
  • DC CURRENTS:
  • Calibres: 50 - 500 μA - 5 - 150 - 500 mA - 1,5 - 10 A
  • Accuracy class: 2
  • Voltage drop: 1,2 V max
  • Protection: 1.6 A and 10 A fuses - 10 kA / 600 Veff
  • AC CURRENTS:
  • Calibres: 500 μA - 5 - 150 - 500 mA - 1,5 - 10 A
  • Accuracy class: 2,5
  • Voltage drop: 1,2 V max
  • Protection: 1.6 A and 10 A fuses - 10 kA / 600 Veff
  • RESISTANCES: 
  • Calibres: x 1 x 10 x 100
  • Middle point: 200 Ω; 2 Ω; 20 Ω
  • Accuracy class: 2,5
  • Protection Electronic: 400 Veff (5 s) + audible alarm
  • CONTINUITY: 
  • Detection threshold: R < 150 Ω
  • OTHER MEASUREMENTS: 
  • Diode test: Yes
  • dB: Yes
  • General specifications:
  • Sealing: IP 65
  • Scale length : 80 mm
  • Safety: IEC EN 61010-1 Edition 2, Cat. III 600 V - Pollution level 2
  • Operating temperature:  -10 °C to +50 °C
  • Power supply: 1.5 V battery (AA, R6)
  • Dimensions / weight: 40 x 98 x 150 mm / 420 g

  • (1) Rugged carry case
  • (1) Test lead set
  • (2) Spare fuses
  • (1) 1.5 V battery (AA, R6) 
  • (1) Instruction Manual