Temperature datalogger compatible with 7 types of thermocouples. EA15 is an EasyView Dual Input Temperature Datalogger capable of-200 Degrees to 1999 Degrees Fahrenheit and -200 Degree to 1360 Degree Celsius measurements. Timer function displays elapsed time plus the time when minimum and maximum readings are taken. Records Max/Min readings for later recall. Built-in datalogger stores up to 8800 data sets. Complete with built-in stand, protective holster, two Type K bead wire temperature probes and batteries.
Type J Temperature Range:-200 to 1994°F (-150 to 1090°C)
Type K Temperature Range: -200 to 1999.9°F (-150 to 1370°C)
Type T Temperature Range: -200 to 752°F (-150 to 400°C)
Type E Temperature Range: -200 to 1598°F (-150 to 870°C)
Type R Temperature Range: 35 to 1999.9°F (2 to 1767°C)
Type S Temperature Range: 35 to 1999.9°F (2 to 1767°C)
Type N Temperature Range: -200 to 1999.9°F (-150 to 1300°C)
NIST Calibration: Not included
Dual input thermometer accepts Type J, K, E, T, R, S, and N thermocouples
Large backlit LCD displays [T1 plus T2] or [T1-T2 plus T1] or [T1-T2 plus T2]
Selectable units of °F, °C, K (Kelvin)
Wide temperature range with 0.1°/1° resolution
Real time clock
Timer function displays elapsed time plus the time when MIN and MAX readings are taken
Data Hold function freezes reading on display
Records Max/Min readings for later recall
Built-in datalogger stores up to 8800 data sets
RS-232 port to transfer data to PC for analysis (PC software and cable included)
Auto Power off to save battery life
Low battery and overrange indication
Calibration Lab
Measurement Ranges:
K - Type : -200.0°F to +1999.9°F (-150.0°C to +1370.0°C)
J - Type : -200.0° F to +1994.0°F (-150.0°C to +1090.0°C)
T - Type : -200.0°F to +752.0°F (-150.0°C to +400.0°C)
E - Type : -200.0°F to +1598.0°F (-150.0°C to +870.0°C)
R - Type : +32°F to +1999.9°F (0.0°C to +1767.0°C)
S - Type : +32°F to +1999.9°F (0.0°C to +1767.0°C)
N - Type : -200.0°F to +1999.9°F (-150.0°C to +1300.0°C)
Resolution:
J, K, T, E, N Type: 0.1°F/°C/K; R, S Type: 1.0°F/°C/K (0.1° for reference only)
Accuracy:
J, K, T, E, N Type: ±(0.05% reading + 1.5°F (0.75°C)
below -148°F (-100°C): add 0.15% of reading for J, K, E and N; 0.45% of reading for T; ±0.05 of reading +2°F(1°C) for R, S
Temperature Coefficient: 0.01% of reading +0.06°F per °F (+0.03 °C) outside the specified +64°F to 82°F (+18°C to 28°C) range
Below -148°F (-100°C): add 0.04% of reading for J, K, E and N type; and 0.08% of reading for T type
Note: Temperature accuracy does not include the accuracy of the probe. Note: The temperature scale is based on the international temperature scale of 1990 (ITS90).
General Specifications:
Display: Dual Display Multi-function LCD with Backlight
Input Protection: 60VDC; 24VAC rms
Measurement Rate: 1 time per 1.5 second
Over range indication: "OL" appears on the LCD
Open input indication: "OL" appears on the LCD
Low battery indication: “BT” appears on the LCD
Manual Memory Capacity: 98 sets
Datalogging Memory Capacity: 8800 sets
Power supply: 6 ‘AAA’ Batteries
Battery life: Approximately 200 hours with alkaline batteries