The HD2114.0, HD2134.0, HD2164.0 micromanometers, the HD2114B.0 barometer, and the corresponding HD2114.2, HD2134.2, HD2164.2 and HD2114B.2 models with datalogger function. They are portable instruments with a large LCD display. They perform measurements of absolute, relative and differential pressure, and temperature.
Pressure is measured using an internal differential module that compares the atmosphere with bottom of a fixed scale. With the PP471 module acting as an interface, the instrument can use all the TP704 and TP705 series Delta Ohm probes to perform the measurements. The HD2114B.0 and HD2114B.2 internal module measures the barometric pressure. The temperature is detected using immersion, penetration, contact or air probes, with SICRAM module or direct 4-wire probes. The sensor can be a Pt100, Pt1000 or Ni1000. The temperature probes are fitted with a SICRAM module, with the factory calibration settings already being memorized inside. On being turned on the instrument automatically detects these settings.
The HD2114.2, HD2134.2, HD2164.2 and HD2114B.2 instruments are dataloggers. They memorize up to 36,000 samples, which can be transferred from the instrument connected to a PC via the RS232C serial port or USB 2.0 port.
The storing interval, printing, and baud rate can be configured using the menu. The HD2114.2, HD2134.2, HD2164.2 and HD2114B.2 instruments can transfer via the RS232C serial port the acquired measurements to a PC or to a portable printer in real time.
The Max, Min and Avg function calculates the maximum, minimum or average values. The Peak function can be activated with external probes connected to the module PP471 and detects the presence of pressure peaks. Other functions include the relative measurement REL, the HOLD function, and the automatic turning off which can also be disabled. The instruments have IP66 protection degree.
Features:
Instrument
Dimensions (Length x Width x Height) 185x90x40mm
Weight 470g (complete with batteries)
Materials ABS, rubber
Display 2x4½ digits plus symbols
Visible area: 52x42mm
Operating conditions
Working temperature -5…50°C
Storage temperature -25…65°C
Working relative humidity 0…90%RH without condensation
Protection degree IP66
Power supply
Batteries 4 1.5V type AA batteries
Autonomy 200 hours with 1800mAh alkaline batteries
Power absorbed with instrument off 20μA
Mains - models HD21...4.2
Output mains adapter 12Vdc / 1000mA
Measuring unit
°C - °F - Pa - hPa - kPa - mbar - bar - atm mmHg - mmH2 O - kgf/cm2 - PSI - inchHg
Security of logged data Unlimited, independent of battery charge conditions
Time
Date and time Real time Accuracy 1min/month max drift