The Druck UNIK 5000 pressure sensing platform is a high performance configurable solution to pressure measurement. The use of Druck silicon technology and analog circuitry enables best in class performance for stability, low power and frequency response. The new platform enables you to easily build up your own sensor to match your own precise needs. This high performance, configurable solution to pressure measurement employs modular design and lean manufacturing techniques.
UNIK 5000 sensors are available with pressure ranges from 1 to 10,000psi. There are three grades of performance (Industrial, Improved and Premium), offering accuracies from 0.2% to 0.04%. A wide range of pressure connectors, electrical connectors, and electrical outputs are available to suit most environments with operating temperature ranges from -67 to 257°F. All units feature high overpressure capability, are constructed from high grade stainless steel and are compliant with global and regional certification bodies (pending).
High levels of performance have been achieved without the use of digital electronics making UNIK5000 sensors particularly well suited to sensitive applications, such as those requiring low power, low noise pulse power and fast response.
UNIK5000 Applications include:
Depth and Level
Industry and process
Oil and Gas
Aerospace
Power Generation
Marine
Custom pressure ranges and cable lengths are available upon request. Please contact us for details.
Ranges from 70 mbar (1 psi) to 700 bar (10000 psi)
Accuracy to ±0.04% Full Scale (FS)
Best Straight Line (BSL)
Stainless Steel construction
Frequency response to 3.5 kHz
High over pressure capability
Hazardous Area certifications
mV, mA, voltage and configurable voltage outputs
Multiple electrical and pressure connector options
Operating temperature ranges from –55 to +125°C (-67 to +257°F)
Online configuration tool
Operating Pressure Ranges:
Gauge Ranges: Any zero based range 70 mbar to 70 bar (1 to 1000 psi)
Note: All psi values are approximate.
Sealed Gauge Ranges: Any zero based range 10 to 700 bar (145 to 10000 psi)
Absolute Ranges: Any zero based range 100 mbar to 700 bar (1.5 to 10000 psi)
Differential Ranges: Wet/Dry: Uni-directional or bi-directional 70 mbar to 35 bar (1 to 500 psi); Wet/Wet: Uni-directional or bi-directional 350 mbar to 35 bar (5 to 500 psi); Line pressure: 70 bar (1000 psi) maximum
Barometric Ranges: Barometric ranges are available with a minimum span of 350 mbar (5.1 psi)
Non-Zero Based Ranges: Non-zero based ranges are available. For non-zero based gauge ranges, please contact Druck to discuss your requirements.
Over Pressure:
10 × FS for ranges up to 150 mbar (2 psi);
6 × FS for ranges up to 700 mbar (10 psi);
2 × FS for barometric ranges;
4 × FS for all other ranges (up to 200 bar for ranges ≤70 bar and up to 1200 bar for ranges >70 bar)*
For differential versions the negative side must not exceed the positive side by more than:
6 × FS for ranges up to 150 mbar (2 psi)
4 × FS for ranges up to 700 mbar (10 psi)
2 × FS for all other ranges up to a maximum of 15 bar (200 psi)
Containment Pressure:
Ranges up to 150 mbar (2 psi) gauge 10 x FS
Ranges up to 70 bar (1000 psi) gauge 6 x FS (200 bar (2900 psi) max)
Ranges up to 70 bar (1000 psi) absolute 200 bar (2900 psi)
Ranges above 70 bar (1000 psi) 1200 bar (17400 psi)*
Differential (-ve port) must not exceed positive port by more than 6 × FS (15 bar (200 psi) maximum.
* 600 bar (8700 psi) for pressure connector options PX, RA and RF